Continuity Function Test; Diode Test Function Test - Fluke ScopeMeter 190-202 Service Manual

Hide thumbs Also See for ScopeMeter 190-202:
Table of Contents

Advertisement

Table 4-18. Meter Volts AC Measurement Verification Points
Range
500.0 mV
500.0 mV
5.000 V
5.000 V
50.00 V
50.00 V
500.0 V
500.0 V
1100 V (1.1 kV)
1000 V

Continuity Function Test

To test the continuity function:
1. Select the Test Tool setup.
METER
2. Press
F1
3. Press
4. Connect the Test Tool to the 5502A as shown in Figure 4-13.
5. Set the 5502A to 20 Ω. Use the 5502A "COMP OFF" mode.
6. Listen to hear that the beeper is on.
7. Set the 5502A to 80 Ω.
8. Listen to hear that the beeper is off.
9. When you are finished, set the 5502A to Standby.

Diode Test Function Test

To do the diode test function:
1. Select the Test Tool setup.
METER
2. Press
F1
3. Press
4. Connect the Test Tool to the 5502A as shown in Figure 4-13.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
5502A output V ac
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
60 Hz
1 kHz
3 kHz
.
to open the Measurement menu and select
.
to open the Measurement menu and select
Performance Verification
Frequency
Meter Reading
494.0 to 506.0
486.0 to 514.0
>350.0
4.940 to 5.060
4.860 to 5.140
>3.500
49.40 to 50.60
48.60 to 51.40
>35.00
494.0 to 506.0
486.0 to 514.0
>350.0
0.980 to 1.020
0.960 to 1.040
> 0.700
Continuity.
Diode.
4
Meter Tests
4-37