Siemens MAMMOMAT 300 Repair Instructions page 17

Hide thumbs Also See for MAMMOMAT 300:
Table of Contents

Advertisement

Removing de cable harness
Connector
X895
X896
X894
X878 *1
X877 *1
X897
X858 *1
X870
X856
X855
X860
4 STOP switch connection
wires
X893 *1
2 protective conductors
X872, X873
X873
X874
5 protective conductors
2 fan connection wires
*1 MAMMOMAT 3000/ 3000 CE only
Siemens AG
Medical Engineering
front of the X-ray tube
7 individual wires, 6 to the X-ray tube (left side) and
1 protective conductor at the collimator.
below the fan
below the fan
below the fan
4 individual wires, 3 at the X-ray tube (right side) and
1 protective conductor at the collimator.
on the front of D805 covered by insulating tubing
on the front of D805 (covered by insulating tubing *1)
on board D805
on board D805
on board D805 (eject wire "2" from the connector)
on the STOP switch
Stereotactic - connection plug on the table support
1 on the X-ray tube (close to the chest wall) and 1 on
the swivel arm frame
on board D805
between board D805 and the compression unit
between board D805 and the compression unit
swivel arm frame
on the fan
Register 5
RXB7-230.091.05
Rev. 03
approximate position
Page 3 of 16
07. 98
TD SD 24
3 - 3
MAMMOMAT 300 / 3000