Removing de cable harness
On the left side of the stand remove:
•
All connectors on board D801 (except for the connection cable to D802).
•
All connectors on board D802
•
Protective conductors above D801 and D802 as well as at the protective conductor bar.
•
Resistor assembly R805 (remove the screws, do not desolder).
•
Wires according to Fig. *1 at connector X880 .
•
Remove all wires at connector X881.
•
Remove relay K801 and remove the socket (cut the rivets).
•
Remove connector X885 by removing the screws*1.
•
Remove connector X801 on display board D803 (front of the stand).
•
Remove connector X886 (inside the stand) *1.
*1 omit for MAMMOMAT 3000 CE.
MAMMOMAT 3000 CE only
•
Remove connector X802 in the base of the stand (transformer T802).
Siemens AG
Medical Engineering
Register 5
RXB7-230.091.05
Rev. 03
Page 9 of 16
07. 98
TD SD 24
3 - 9
3
3
MAMMOMAT 300 / 3000