Removing de cable harness

On the left side of the stand remove:

All connectors on board D801 (except for the connection cable to D802).
All connectors on board D802
Protective conductors above D801 and D802 as well as at the protective conductor bar.
Resistor assembly R805 (remove the screws, do not desolder).
Wires according to Fig. *1 at connector X880 .
Remove all wires at connector X881.
Remove relay K801 and remove the socket (cut the rivets).
Remove connector X885 by removing the screws*1.
Remove connector X801 on display board D803 (front of the stand).
Remove connector X886 (inside the stand) *1.
*1 omit for MAMMOMAT 3000 CE.

MAMMOMAT 3000 CE only

Remove connector X802 in the base of the stand (transformer T802).
Siemens AG
Medical Engineering
Register 5
RXB7-230.091.05
Rev. 03
Page 9 of 16
07. 98
TD SD 24
3 - 9
3
3