Installing the cable harness

MAMMOMAT 300/ CE

Insert the cable harness in the swivel arm.
Reattach the table support with the screws.
NOTICE
4 emergency
stop button
cables (backside)
Fig. 9
Siemens AG
Medical Engineering
Make sure that the shielding is folded over the lead-through ring
and that the semicircular cable clamp is in position. Also make
sure that the braided grounding wire is between the hose clamp
and the shielding.
X896
LEFT SIDE
Register 5
X895
X894
RXB7-230.091.05
Rev. 03
07. 98
X897
RIGHT SIDE
Page 7 of 14
MAMMOMAT 300 / 3000
TD SD 24
4 - 7
4
X855
X856
X860
X870