Preliminary work
Fig. 7
Position of the anchor plate
Remove the anchor plate by removing the screws (1/Fig. 7).
2
Fig. 8
Components on object table

MAMMOMAT 3000 / MAMMOMAT 3000 CE only

Perform the following work on the table support for WING 1:
Remove cover (2/Fig. 8) by removing two screws on the back.
Remove the screws holding the connector X893 (4/Fig. 8). Mark the top of the
connector.
Remove the screws holding the connector X891 (6/Fig. 8). Mark the top of the
connector.
Remove the screws holding the assembly plate (3/Fig. 8).
Remove the table support cover (5/Fig. 8).
Loosen bellow (7/Fig. 8) by removing plastic guide blocks inside the bellow.