OVERVIEW

The MLX light source delivers 300 spare watts of cool white infrared (IR) filtered light. The lamp is housed in a lamp
module that can be easily serviced without special tools.
set up and inspection Before use
the light source comes with the hospital grade power cord packaged separately. Verify that both components are in good
condition.
Before turning power on to the light source, make sure the unit is plugged into any standard 100Vac to 240Vac 47-
63hz (as appropriate) outlet. grounding reliability is guaranteed only when connected to a "hospital grade" receptacle.
Allow a minimum of 2 inches (5.08 cm) clearance at the rear and sides of the unit for cooling air flow. Ensure that the
unit is not near air exhaust or against other equipment.
warning: the MLX light source monitors air intake temperature and will turn off if ambient temperature is > 40

OPERATION

1. Insert a fiber optic cable in the proper port in the turret. The active port is indicated by a on
(see Figure 1). available ports are acMi, olympus, storz and wolf cable-compatible. all ports
appropriately.
2. Press the Power switch in the upper left corner of the front panel (see Figure 2).
3. when the Power switch is pressed the stand-by Light (Figure 2) will blink and there will be a 3 – 4 second delay
before the Xenon Lamp illuminates while the system does a self-diagnostic check.
4. Press the stand-by Button.
active Point
Figure 1
M L X 3 0 0 w a t t X e n o n L i g h t s o u r c e
stand-by Button
c
o
the front panel
are identified
Power switch
Figure 2
5