M L X 3 0 0 w a t t X e n o n L i g h t s o u r c e
XENON LAMP MODULE REPLACEMENT
to remove:
1. Make sure the power is off and the hospital grade power cord is disconnected.
2. remove the retaining screws from the top plate (Figure 4) and retain screws.
3. slide the top plate towards the back of the unit until it stops.
4. Lift the top cover up and off the unit.
5. unscrew the lamp module door and remove (Figure 5) and retain screw.
6. grasp the module by the tabs and pull directly up (Figure 6).
Figure 4
Figure 5
reset
Figure 6
Figure 7
to replace:
1. orient the lamp module (P/n 001320LX) to face forward.
2. Place the module into the light source.
3. seat module securely.
4. replace the lamp module door and secure it with the screw.
5. align top cover brackets into slots and slide forward.
6. slide the top plate forward until it engages the front panel.
7. replace the retaining screws in the rear panel.
8. replace the hospital grade power cord and turn on power to verify operation.
to reset Lamp age Meter:
1. using the membrane switch, turn the display to the lamp age/system age screen.
2. on the rear of the unit insert a pen/pencil or other suitable pointed object into the slot marked reset (Figure 7).
3. Push until a click is felt.
4. Verify that the lamp age has been reset to 0.
note: the system age reading cannot be reset.