Section 5
Installation
Install Control Panel
in Remote Loc. (cont.)
5-4
Nanopure™ Analytical
4. Using the wall bracket as a template, locate and drill the mounting
holes in the wall. A minimum of two (customer-supplied) fasteners will
be required. Attach the wall bracket to the wall using the customer-
supplied fasteners.
5. Locate the 25-pin connector on the back top right of the unit. Remove
the protective black cover and store it for future use if the control panel
is returned to the top cover.
6. Locate the 10 ft. (3.1 m.) 25 pin M-F cable included with the unit;
attach the female end to the 25 pin connector on the back top right of
the unit, and the other end to the connector on the bottom of the
control panel.
7. Tighten the cable screws on each end to secure the cable to the unit
and control panel.
8. The control panel may now be mounted on the wall.
Caution Repeated removal and replacement of the control panel from the
top cover may eventually cause it to become damaged.
Note A bracket is available that will allow you to mount the control panel
on the bench. If bench mounting is desired, order bracket AY1367X2.
Wall Mount
Bracket
Blank
panel
Figure 5-1. Control Panel Wall Mounting