MAIN PCB REPLACEMENT

If an 825D main board is replaced, the 825D will boot up to this screen:
The system will also check whether the 825D's main board has been replaced, so the scale
configuration can be downloaded from the controller to the main board.
If the operator selects YES, then the scale configuration will be downloaded from the controller
to the mainboard. Scale configuration includes the number of load cells, all load cell IDs, and
individual load cell trim. The 825D indicator parameters will need to be entered manually
(Interval, Decimal Point Position, Zero Tracking, Filtering, Print Settings, and Serial Settings).
NOTE: A dead load calibration will need to be performed (does not require test weights).
69
8545-0980-0M Rev C • 825D Weight Digital Indicator