6. Maintenance

1. Check the power cable and power supply for damage and scratches.
2. Check the mattress cover to make sure it is undamaged and that covers and cells are
correctly assembled.
3. Check that air is coming from the outlets on the side of the pump.
4. Check that no tubes or connectors are damaged or pinched.
Contact
the manufacturer or
your local distributor for spare parts.
6.1 Air filter and fuse
replacement
Air filter (A):
1. Loosen the small protective plate on
the rear of the pump using a
screwdriver.
2. Remove the filter from the holder.
3. Insert the new filter in the holder,
put the protective plate back in place
and secure using the screws.
A
Fuses (B):
1. Use a screwdriver to remove the
fuse holder.
2. Replace the old fuse with a new one
marked T1A/250 V; note that there are
two fuses.
3. Use a screwdriver to put the fuse
holder back in place.
B