Chapter 5 Troubleshooting
82
Flexibility adjustment
(for endoscopes with flexibility adjustment only)
Irregularity
description
Too difficult to
The insertion tube is
turn the
looped.
flexibility
adjustment ring.
Auxiliary water feeding
(for endoscopes with auxiliary water feeding only)
Irregularity
description
The auxiliary
The auxiliary water
water inlet cap
inlet cap is worn out.
is leaking.
The auxiliary water
inlet cap is incorrectly
installed.
Endo-therapy accessories
Irregularity
description
Endo-therapy
An incompatible
accessory does
endo-therapy
not pass
accessory is being
through the
used.
instrument
channel
smoothly.
Others
Irregularity
description
The remote
The wrong remote
switch does not
switch is operated.
work.
The remote switch
function has been set
improperly.
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
Possible cause
Straighten the insertion tube.
Possible cause
Replace it with a new one.
Install the auxiliary water inlet cap correctly.
Possible cause
Refer to the "System chart" in the Appendix
and select a compatible endo-therapy
accessory. Confirm that the color code on
the endo-therapy accessory matches that
on the endoscope.
Possible cause
Operate the correct remote switch.
Set the remote switch function correctly as
described in the video system center's
instruction manual.
Solution
Solution
Solution
Solution