Chapter 3 Preparation and Inspection
42
Inspection of the UP/DOWN angulation mechanism
1.
Move the UP/DOWN angulation lock all the way in the opposite direction of
the "F
" mark. Then turn the UP/DOWN angulation control knob in the
"
U" or the "D
2.
Confirm that the angle of the bending section is roughly stabilized when the
UP/DOWN angulation control knob is released.
3.
Confirm that the bending section straightens out when the UP/DOWN
angulation lock is moved all the way in the "F
UP/DOWN angulation control knob is released.
Inspection of the RIGHT/LEFT angulation mechanism
(except GIF-N180)
1.
Turn the RIGHT/LEFT angulation lock all the way in the opposite direction of
the "F
" mark. Then turn the RIGHT/LEFT angulation control knob in the
"R
" or the "
2.
Confirm that the angle of the bending section is roughly stabilized when the
RIGHT/LEFT angulation control knob is released.
3.
Confirm that the bending section straightens out when the RIGHT/LEFT
angulation lock is turned in the "F
angulation control knob is released.
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
" direction until it stops.
L" direction until it stops.
" direction and the RIGHT/LEFT
" direction and the