Chapter 2 Instrument Nomenclature and Specifications
2.4
Attaching the chain for water-resistant cap
(MAJ-1119)
32
Do not lift the endoscope by the chain for water-resistant cap.
Otherwise, operator and/or patient injury can result, or the
endoscope and/or water-resistant cap may be damaged
when the fitting comes off the S-cord connector mount or the
chain connector of the GIF-N180.
Only connect the fitting to the S-cord connector mount or the
chain connector of the GIF-N180. Connecting the fitting to
the suction connector may impair the connection of the
suction tube to the suction connector. It may also cause the
suction tube to become disconnected from the endoscope
and allow patient debris to spray.
The chain for water-resistant cap and water-resistant cap
itself cannot be ultrasonically cleaned; doing so could
damage them. The water-resistant cap with the chain can
only be ultrasonically cleaned if connected to endoscopes
that are being cleaned in an endoscope reprocessor (such as
OER) with an ultrasonic cleaning phase.
When attaching the water-resistant cap to the electrical
connector, do not pinch the chain for water-resistant cap
between the electrical connector of the endoscope and the
water-resistant cap. Otherwise, equipment damage may
result.
The chain for water-resistant cap and water-resistant cap
cannot be ETO gas sterilized; doing so may damage them. If
the water-resistant cap is connected to the endoscope by the
chain, be sure to remove the chain and the water-resistant
cap from the endoscope before ETO gas sterilization.
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL