Chapter 4 Operation
Transporting outside the hospital
78
Transport the endoscope in the carrying case.
Always clean, disinfect or sterilize the endoscope after
removing it from the carrying case. If the endoscope is not
cleaned, disinfected or sterilized, it could pose an
infection-control risk.
The carrying case cannot be cleaned, disinfected or
sterilized. Clean and disinfect or sterilize the endoscope
before placing it in the carrying case.
Do not attach the water-resistant cap when transporting the
endoscope, to avoid damage to the endoscope caused by
changes in air pressure.
Before putting the endoscope in the carrying case, always
make sure that the insertion tube is set to the most-flexible
condition. Putting the endoscope in the carrying case while
the insertion tube is rigid could damage the endoscope (for
endoscopes with flexibility adjustment only).
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL