Immunity test
Conducted RF
3 Vrms
(150 kHz – 80 MHz)
IEC 61000-4-6
Radiated RF
3 V/m
(80 MHz – 2.5 GHz)
IEC 61000-4-3
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications
equipment such as cellular phones
IEC 60601-1-2
Compliance
test level
3 V (V
3 V/m (E
•
Where "P" is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and "d" is the recommended separation
distance in meters (m).
•
This instrument complies with the requirements of
IEC 60601-1-2: 2001. However, under electromagnetic
environment that exceeds its noise level, electromagnetic
interference may occur on this instrument.
•
Electromagnetic interference may occur on this instrument
near a high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
level
Formula for recommended separation distance
(V
=E
=3 according to the compliance level)
1
1
)
3.5
1
d
=
-------- -
V
1
)
3.5
1
d
=
-------- -
E
1
7
d
=
---- -
E
1
Guidance
P
P
80 MHz – 800 MHz
P
800 MHz – 2.5 GHz
Appendix
103