Chapter 3 Preparation and Inspection
Inspection of the auxiliary water tube
(for endoscopes with auxiliary water feeding only)
Inspection of the mouthpiece (for GIF models only)
46
Inspect the auxiliary water tube for cracks, scratches, flaws and other damage
(see Figure 3.13).
Figure 3.13
Do not use a mouthpiece that is damaged, deformed or
reveals other irregularities. Doing so may cause patient injury
and/or equipment damage.
Placing the mouthpiece in the patient's mouth before the
procedure prevents the patient from biting and/or damaging
the endoscope's insertion tube.
1.
Confirm that the mouthpiece is free from cracks, deformation or
discoloration (see Figure 3.14).
2.
Using your fingers, check for excessive scratching or other irregularities on
all surfaces of the mouthpiece (see Figure 3.14).
EVIS EXERA II GIF/CF/PCF TYPE 180 Series OPERATION MANUAL