5.

Turning OFF the High-Voltage Switch

1Y
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
Turn OFF the high-voltage switch (S1) of the SCN08D board.
009-058-00
10.20.2000 FM2732 (2)
At intervals of one year, or when 60,000 IPs have been processed
Turn OFF.
SCN08D board
CR-IR347 Service Manual
HHS label #2
Scanner unit
OFF position
PM - 20
High-voltage
switch