Cleaning Of The Electrode; Electrode Lifetime; Electrode Spare Parts - Xylem mjk pHix Compact Manual

Hide thumbs Also See for mjk pHix Compact:
Table of Contents

Advertisement

Cleaning of the electrode

Use clean water for cleaning the electrode.
A 10% hydrochloric acid solution can be used for cleaning if the electrodes are
greasy. Protein coatings are removed with Pepsin cleaning fluid.

Electrode lifetime

The expected lifetime of the electrodes is dependent of the measurement. The
lifetime is given with reservation as the physical conditions as well as the
temperature, pressure and pollution of the fluid has great influence on the
stated lifetime.
The lifetime is based on 77° F / 25° C, but if the temperature is doubled, the
lifetime is halved!
pH electrodes in purifying plants for ordinary household waste-water lasts
approximately one year, whereas the lifetime of pH electrodes in industrial
plants is dependent on measuring application and the process. Redox
electrodes have a lifetime of approximately two years, depending on cleaning.

Electrode spare parts

Item
Description
no.
160310 pHix® pH electrode, 0-point=4,6 pH
160311 pHix® pH electrode, 0-point=7,0 pH
160312 pHix® redox electrode
571030 Black rubber gasket for electrode
521409 Union ring w. Viton gasket for mounting of pHix® on pipe - NOT
AVAILABLE IN THE US
521441 Viton gasket for union ring - NOT AVAILABLE IN THE US
Maintenance / Cleaning of the electrode
pHix® Compact
EN 4.10  Version 2007
32