Peak Scientific MS Table Series User Manual
Hide thumbs Also See for MS Table Series:

Advertisement

Quick Links

MS Table (All Models)
User Manual
Copyright © Peak Scientific Instruments Ltd 2017 — MS Table User Manual | Rev 2 | PN UM-0012| RSID 1927 | EN —25/08/2017

Advertisement

Table of Contents
loading

Summary of Contents for Peak Scientific MS Table Series

  • Page 1 MS Table (All Models) User Manual Copyright © Peak Scientific Instruments Ltd 2017 — MS Table User Manual | Rev 2 | PN UM-0012| RSID 1927 | EN —25/08/2017...
  • Page 2: Table Of Contents

    Contents Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration of Conformity Environmental Declaration Technical Specification Technical Specification Fittings Kit Contents Unpacking Instructions Installation Generator Environment Generator Overview 1.5L & 3.5L General Dimensions Rear Connections Drain Connection Electrical Connection...
  • Page 3: Change History

    Users can refer to the contents page to find the relevant information. Please review each of the following sections carefully. Thank you for selecting Peak Scientific to meet your gas generation needs, and should you require any further assistance or support please do not hesitate to contact Peak Scientific or the Peak Partner from which you purchased your generator.
  • Page 4: Introduction

    Introduction The MS table has been designed as a Sciex exclusive product to house the Triple Quad™, QTRAP™ or IVD MS systems. The MS table comes in a standard and premium form, the premium version including an inbuilt Infinity 1031 allowing the supply of curtain, source and exhaust gases.
  • Page 5: Warranties And Liabilities

    Warranties and Liabilities 1. The Company warrants that it has title to the Goods. 2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch...
  • Page 6: Safety Notices

    If the user employs an operating procedure, item of equipment or a method of working which is not specifically recommended by Peak Scientific, the user must ensure that the equipment will not be damaged or become hazardous to persons or property.
  • Page 7: Declaration Of Conformity

    Declaration of Conformity Peak Scientific Instruments Ltd. Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE Declare that: Equipment: Nitrogen Gas Generator Models: MS Table To which this declaration relates, is in conformity with the applicable EC directives, harmonized standards, and other normative requirements.
  • Page 8: Environmental Declaration

    WEEE (Waste of Electrical and Electronic Equipment) • 2011/65/EU RoHS 2 (Restriction of Hazardous Substance) Peak Scientific Instruments Ltd fully complies with its obligations towards the European WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These obligations are being met within the B2B compliance group.
  • Page 9: Technical Specification

    Technical Specification Environment Minimum Operating Ambient Temperature 5°C (41°F) Maximum Operating Ambient Temperature 30°C (86°F) Maximum Altitude 2000 meters Maximum Relative Humidity 70% Non-Condensing Minimum Storage Temperature* -20°C (-4°F) Maximum Storage Temperature* 60°C (140°F) Inlet Conditions Maximum Inlet Air Pressure 8.3-10 bar (120-145 psi) Minimum Air Inlet Flow 115 L/min (4.06 cfm)
  • Page 10: Technical Specification

    Technical Specification Environment 2N Hi-Flow Minimum Operating Ambient Temperature 5°C (41°F) Maximum Operating Ambient Temperature 30°C (86°F) Maximum Altitude 2000 meters Maximum Relative Humidity 70% Non-Condensing Minimum Storage Temperature* -20°C (-4°F) Maximum Storage Temperature* 60°C (140°F) Inlet Conditions Maximum Inlet Air Pressure 8.3-10 bar (120-145 psi) Minimum Air Inlet Flow 115 L/min (4.06 cfm)
  • Page 11 Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to unpacking. Any damage should be reported immediately to the carrier and Peak Scientific or the Peak Partner from where the unit was purchased. Follow the unpacking instructions posted on the side of the crate. It will require two people to remove the unit from the shipping crate and to manoeuvre the air compressor onto the bench.
  • Page 12: Fittings Kit Contents

    Fittings Kit Contents Supplied in the Fittings Kit are all the fittings required to connect the generator to the application. The contents of the Fittings Kit are as follows: 1. ¼” Teflon Tubing* x 12m 2. 6mm PE Tubing* x 3m 3.
  • Page 13: Installation

    Installation Generator Environment The generator is designed for indoor use only. It should be installed adjacent to the application(s) it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe.
  • Page 14: Generator Overview

    Generator Overview General Dimensions 920.00 mm 800.00 mm The generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the generator. WARNING Page 14...
  • Page 15: Rear Connections

    Rear Connections Supplied in the Fittings Kit are all the fittings required to connect the MS Table Generator to the application. The contents of the Fittings Kit are as follows: Exhaust Output Source Output Curtain Output Air Input Drain Power Inlet Mains Power Cable Drain Connection For users of the 1N and 2N Hi-Flow, use the ¼”...
  • Page 16: Electrical Connection

    Electrical Connection Connect the Generator to an appropriate 110 or 230 volt single-phase supply, refer to the generator serial plate for input specification and ensure your supply matches the requirements. If the appropriate power cord is not supplied; a new plug, rated to at least 13 amps, can be fitted by a qualified electrician.
  • Page 17: Air Connection

    A minimum inlet pressure of 120 psig (8.3 barg) is required. Any doubts as to the suitability of your compressed air supply should be referred to Peak Scientific or any of their authorised partners. The Compressed Air supply should be connected to the inlet located at the rear of the unit as shown in the Rear Connections section of this manual.
  • Page 18: Connecting To The Application

    ⁄ ” I/D). Tubing and fittings not supplied in the fittings kit. > 40 metres: Please contact Peak Scientific with the relevant distance and we will calculate the flow resistance and the tubing size required. A combination of ¼”/ ⁄...
  • Page 19: Normal Operation

    Unusual Operation If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact Peak Scientific or the Peak Partner from which the generator has been purchased. Relocation of MS Table 1.
  • Page 20: Safe Operation Of Pump Tray

    Safe Operation of Pump Tray For safe installation or removal of roughing pumps, we recommend use of the sliding function of the pump tray, as shown below. This should only be attempted once the Mass Spec has been positioned on top of the table. WARNING Remove the front panel, ensuring that the fan power and earth cables are disconnected.
  • Page 21: Infinity 1031 Installation

    Zero Grade air. The MS Table 1B does not have an integrated Nitrogen Generator, but does have the facility to bolt-on to the side of the table a Peak Scientific Instruments Wall Mounted Infinity 1031 Nitrogen Generator. Using the remaining M5 fittings from the fittings kit, mount the Infinity 1031 to the side of the MS Table 1B.
  • Page 22: Service Requirements

    Service Requirements Service Schedule Purchase Interval Component Visit Genius MS Table 1N Annual Maintenance Genius MS Table 2N Hi-Flow Annual 12 Months www.peakscientific.com/ordering Maintenance Kit Infinity 1031 Annual Maintenance Kit Page 22...
  • Page 23: Peak Protected

    Peak Protected With Peak Scientific you invest in not only a product but peace of mind. With a network of certified Peak engineers stationed throughout the globe, Peak’s rapid response team are never far away and our commitment is to keep your generator running day in, day out, protecting your laboratory workflow.
  • Page 24: Cleaning

    Cleaning Clean the outside of the generator only using warm soapy water and a clean damp cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use. Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the Generator.
  • Page 25: Troubleshooting

    Troubleshooting Problem Possible Solution • Contact your service provider. The green LED has not illuminated but the power switch is illuminated. • Check for leaks between the generator and instrument. The instrument is reporting low pressure. • Contact your service provider. The unit is switched on but the fans are not •...
  • Page 26 See our enclosed Peak [Protected] leaflet for further information. Important! You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
  • Page 27 Fax: +44 141 812 8200 For further information on any of our generator products please contact marketing@peakscientific.com Copyright © Peak Scientific Instruments Ltd 2017 — MS Table User Manual | Rev 2 | PN UM-0012| RSID 1927 | EN —25/08/2017...

This manual is also suitable for:

1b1n2n hi-flow

Table of Contents