Transport within the hospital
Transport refers to any movement of the medical
device without the patient that does not serve to po-
sition the medical device.
Increasing the tipping stability
Swivel the control and display unit (Medical
Cockpit) until it is centrally aligned with the ven-
tilation unit.
Set the hinged arms to minimum extension.
Drain the water container of the breathing gas
humidifier.
Secure the breathing gas humidifier to the trol-
ley, not to the side rails of the ventilation unit.
Do not attach any additional parts to the side
rails of the ventilation unit.
If fitted, slide the bed coupling into its retracted
position.
Grasp the trolley handle firmly and push the de-
vice in longitudinal direction.
Safety information for patient transport within the
hospital also applies, see instructions for use.
Assembly and preparation
11