Dräger Infinity Acute Care System Workstation Critical Care Supplement To Instructions For Use page 26

Table of Contents

Advertisement

Alarm – Cause – Remedy
Alarm
Alarm message
priority
!
200
Setting not confirmed
!!!
255
Standby mode activated
!
200
Ventilation mode not
confirmed
!!
140
VThf not reached
Applies only to
Babylog VN500.
26
Cause
One or more settings have
been changed but not
confirmed.
Device has been switched
to standby mode.
The ventilation mode has
been changed but not
confirmed.
Tidal volume could not be
reached due to increased
resistance.
Tidal volume could not be
reached due to pressure
limitation of the device.
Remedy
If necessary, change these
settings and confirm the
change with the rotary
knob.
Acknowledge standby
mode by touching "ALARM
RESET" button and confirm
with rotary knob.
If necessary, change the
ventilation mode and
confirm the change with the
rotary knob.
Check patient condition.
Check ventilation settings.
Reduce fhf or increase
I:Ehf to 1:1.