Set a SID o 100 cm (39"). Use the measuring tape.
Switch on the main power switch.
Switch on the light localizer and align the cassette. Now, collimate the light field to
25 x 25 cm (10" x 10").
Mark the four corners with coins and mark one of the corners with two coins (Fig. 12).
Set 52 kV, 2.0 mAs on the control panel.
CAUTION
Release an exposure.
Develop the film and measure the deviation. Acceptable deviation ± 1.7% (Fig. 13).
E
E
Fig. 14
If the deviation is greater, loosen the Allen screws marked "E" slightly and move the
rotary-flange to desired position.
Tighten the screws marked "E" (Fig. 14).
Repeat the test and adjustment of the rotary flange position until the deviation between
the light and radiation fields is within the tolerance specified.
Reinstall covers "A" and "B".
MOBILETT Plus / - E/ - M