Siemens MOBILETT Plus Service Instructions Manual page 18

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 8
Dis-
played
messages
Err 16
Err 17
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus / - E/ - M
Mode of
Cause
operation
Generator ON
The capacitors are
charging too rapidly
(>7 V/s)
Generator ON
The capacitors are
charging too slowly
(< 1.4 V/s)
SPR8-215.898.01
Rev. 05
Display messages
Possible reason
Check the capacitance in the
capacitor bank, refer to section
"P03 - Capacitor bank and
charging test".
If the capacitance is too low, one or
more capacitor rows are
disconnected. If there is an internal
error in a capacitor, the whole row
is disconnected by a fuse, refer to
page 130** and section "Capacitor
bank".
If the error appears after working
on the capacitor bank, it may be
that the internal residual voltage in
the modules was too high when
they were connected to each other.
This caused the fuse to blow, refer
to page 130**.
Check that S10 on D7 is switched
off.
If the error appears after working
on the capacitor bank, check the
cabling and the connections to the
capacitor bank, refer to page
130**.
Check fuse U8 on D8 and fuse U1
on D7.
Check the capacitance in the
capacitor bank, refer to section
"P03 - Capacitor bank and
charging test".
Internal short-circuiting in a
capacitor or in the capacitor bank,
refer to section "Capacitor bank".
Page 8 of 22
01.02
SPS-UD
Siemens-Elema AB
Solna, Sweden