Siemens MOBILETT Plus Service Instructions Manual page 25

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Display messages
Dis-
played
messages
CAL 28
CAL 29
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1
Siemens-Elema AB
Solna, Sweden
Mode of
Cause
operation
Generator ON
An error was
during start-up
detected during pro-
gram execution. The
program is inter-
rupted.
Generator ON
The single tank is
not correctly con-
nected.
SPR8-215.898.01
Rev. 05
Possible reason
Check that all supply voltages on
D1 are present, refer to page 110**.
If CAL 28 occurs again or remains
displayed: the ROM containing the
program may be defective.
Replace the software (IC's I72, I66
and I74) on D1 ; refer to page
110**.
If the error remains after replacing
the software, D1 should be
replaced.
Enter the cause and the error
message "CAL 28" in the service
protocol.
The single tank is missing or the
cable harness is incorrectly
connected.
Check that connection K9 on D1 is
connected.
Check for possible damage in the
main cable harness between D1
and the single tank. Measure K9:
13-4 and move the arm and single
tank while the connections are
being measured. See page 140**.
If damage is detected, replace the
main cable harness (both cables).
Refer to the section "Replacement
of the arm cable harness".
Page 15 of 22
01.02
SPS-UD
2 - 15
MOBILETT Plus / - E/ - M

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus m30401081020001

Table of Contents