Siemens MOBILETT Plus Service Instructions Manual page 76

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

9 - 6
Replace the cable ties (Fig. 8). The cable ties have to be mounted exactly as shown in
Fig. 9 and Fig. 10. The cable ties should not be tightened hard. Fig. 9 and Fig. 10 shows
how the cable tie on the right side is mounted. Repeat for the left side, but with the cable
tie mirror-inverted.
NOTE
Fig. 9
Rotate the collimator through the whole interval, and make sure the cable ties does not
come in contact with the collimator / moving parts.
Reattach collimator cover "C".
NOTICE
MOBILETT Plus / - E/ - M
Replacement and repairs on the collimator
Make sure not to forget to attach the cable ties!
MOB005
When replacing the collimator, the new collimator should be
positioned and adjusted so that the light and radiation fields
correspond.
The adjustment should be performed by moving the whole
collimator, not just changing the position of the collimator bulb.
SPR8-215.898.01
Rev. 05
MOB00579
Fig. 10
Page 6 of 8
01.02
SPS-UD
Siemens-Elema AB
Solna, Sweden

Hide quick links: