Safety Information And Protective Measures - Siemens MOBILETT Plus Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

General

Safety information and protective measures

CAUTION
WARNUNG
WARNING
Siemens-Elema AB
Solna, Sweden
When performing service work and tests adhere to:
- the product-specific safety information in the documents,
- as well as the general safety information contained in ARTD
Part 2.
Connect the MOBILETT Plus/Plus E/Plus M only to mains power
supply outlet (receptacle) , that corresponds to the installation
requirements of VDE 0107 or country-specific regulations.
Remove or install boards only when the generator is switched
OFF. Adhere to the ESD guidelines
Checks and adjustments performed with radiation ON are
identified by the radiation warning symbol
types of adjustments, radiation protective clothing must be
worn.
Never work with the system open if the capacitor is charging or
has just charged.
The capacitor bank may still be charged even if the system is
switched OFF and the mains cable is disconnected. Life-
threatening electric shock hazard exists.
To avoid danger, refer to the section entitled
"Protective measures when working on the capacitor bank".
SPR8-215.898.01
Rev. 05
.
Page 3 of 6
01.02
SPS-UD
1 - 3
1
. During these
MOBILETT Plus / - E/ - M

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus m30401081020001

Table of Contents