Power On/Off And Power On Tests; Reset; [Keys] - Stanford Research Systems SR830 Manual

Dsp lock-in amplifier
Table of Contents

Advertisement

Signal Inputs
Power

Reset

[Keys]

CH1 Display
The power switch is on the rear panel. The SR830 is turned on by push-
ing the switch up. The serial number (5 digits) is shown in the CH1 and
CH2 displays and the firmware version is shown in the Ref display at
power on.
A series of internal tests are performed at this point.
DATA
Performs a read/write test to the processor RAM.
BATT
The nonvolatile backup memory is tested. Instrument settings are stored
in nonvolatile memory and are retained when the power is turned off.
PROG
Checks the processor ROM.
DSP
Checks the digital signal processor (DSP).
rCAL
If the backup memory check passes, then the instrument returns to the
settings in effect when the power was last turned off (User). If there is a
memory error, then the stored settings are lost and the standard (Std)
settings are used.
To reset the unit, hold down the [Setup] key while the power is turned on.
The unit will use the standard settings. The standard setup is listed on
the next page.
The keys are grouped and labelled according to function. This manual
will refer to a key with brackets such as [Key]. A complete description of
the keys follows in this section.
CH2 Display
Analog Outputs
4-1
FRONT PANEL
Ref Display
Ref Input
Sine Output

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents