Menu item: "CUSTOMIZE":
Here, you have the option of performing your own material calibrations or of customizing existing
calibrations to suit your particular needs. Two options are available for this:
1-point calibration:
This material calibration option merely adjusts (offsets) the configured calibration. As the slope is not
changed, it is important at the start to select a calibration curve that suits the material.
NOTE:
You require a sample of the material being measured to perform a 1-point material
calibration. The moisture value must be determined with another method (weighing and
drying, or similar) before the calibration.
Procedure:
Page 18
1-point calibration:
- Adjusts the calibration curve to the selected point.
- The slope is not changed.
- Just one material sample is required.
2-point calibration:
- Creation of a linear calibration between two measured
points
- Two material samples with different moisture values are
required
At the start of the calibration, the calibration memory
(01 – 15) to be overwritten must be selected with the
"Up"
and "Down"
accept the setting.
Then set the percentage moisture using the "Up"
"Down"
keys, and again press the "Measure" key
confirm your entries.
MMP20 SONO Mobil HD2 (option E)
keys. Press the "Measure" key
Endress+Hauser
to
and
to