Endress+Hauser MMP20 SONO Mobil HD2 Operating Instructions Manual page 19

Moisture measurement, option e
Table of Contents

Advertisement

MMP20 SONO Mobil HD2 (option E)
CAUTION:
If you select "SAVE" at the end of the calibration, you overwrite one of the pre-configured
calibrations (or one you may have already modified) in the probe! The original
calibrations can only be restored with a PC and the Pico-Config software, as well as an
RS485 adapter (e.g. SM-USB).
Endress+Hauser
To start measuring the material, press the "Measure" key
again. Four measurements are taken to increase the accu-
racy. An average of these measurement values is then cal-
culated. The measuring time is approximately 20 seconds.
When the measurement is finished, the measured pulse
transit time is displayed briefly.
NOTE:
Before you start measuring, make sure that the
rods of the probe are fully immersed in the mate-
rial to be measured. The probe must remain in the
material for the entire duration of the measure-
ment and should not be moved.
After this, the calibration can be saved to the calibration
memory set at the start ("Save"). The selected memory is
only overwritten once the "Measure" key
word "OWN:" now appears in front of the original memory
name to clearly indicate which memory has been overwrit-
ten.
is pressed. The
Page 19

Advertisement

Table of Contents
loading

Table of Contents