Summary of Contents for Endress+Hauser Solitrend MMP40
Page 1
These Brief Operating Instructions are not a substitute for the Operating Instructions pertaining to the device. Detailed information is provided in the Operating Instructions and other documentation. Available for all device versions via: • Internet: www.endress.com/deviceviewer • Smartphone/tablet: Endress+Hauser Operations app...
Associated documentation Solitrend MMP40 Associated documentation Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 About this document Symbols used 2.1.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury.
Solitrend MMP40 Basic safety instructions WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury.
Basic safety instructions Solitrend MMP40 ‣ Be familiar with federal/national regulations. ‣ Before starting work: personnel must read and understand the instructions in the manual and supplementary documentation as well as the certificates (depending on the application). ‣ Personnel must follow instructions and comply with general policies.
Solitrend MMP40 Basic safety instructions Workplace safety For work on and with the device: ‣ Wear the required personal protective equipment according to federal/national regulations. Operational safety Risk of injury! ‣ Operate the device only if it is in proper technical condition, free from errors and faults.
Product description Solitrend MMP40 Product description TDR material moisture sensors for measuring bulk solids and media with higher material densities and conductivity values up to 20 mS/cm. Product design A0040142 1 Product design Measuring cell; ceramic (silicon nitride) Sensor plate...
‣ Enter the serial number from the nameplate into the Endress+Hauser Operations App or scan the 2-D matrix code on the nameplate. All of the information on the measuring device and on the scope of the technical documentation pertaining to the device is displayed.
Mounting Solitrend MMP40 Mounting Mounting requirements • The device must be installed at a point in the process in such a way that ensures a constant bulk density, as the bulk density directly affects the calculation of the water content. Where...
Page 9
Solitrend MMP40 Mounting 132 (5.2) 10 (0.39) 86 (3.39) 6.5 (0.26) A0037423 4 Mounting flange for round sensor, short version or round sensor, middle version. Unit of measurement mm (in) The mounting flange serves as a template for the mounting bore holes and the cut-out for the...
Page 10
Mounting Solitrend MMP40 Round sensor, long The round sensor, long version, can be mounted with a mounting frame. A0046911 6 Round sensor mounted, long version, rear view Vessel wall Mounting frame with clamping ring Round sensor, long The mounting frame for the round sensor, long version, can be welded onto the floor or side wall of the vessel.
Page 11
Solitrend MMP40 Mounting 70 (2.76) A0044664 7 Mounting frame for round sensor, long version. Unit of measurement mm (in) Before welding the mounting flange at the mounting location: Check the fit between the sensor and mounting frame Make the cut-out for the sensor at the mounting location Mount the sensor and align it ...
Page 12
Mounting Solitrend MMP40 7.5 (0.3) 90 (3.54) 15 (0.59) 25 (0.98) 120 (4.72) 135 (5.31) 20 (0.79) A0037426 8 Dimensions. Unit of measurement mm (in) Endress+Hauser...
Page 13
Solitrend MMP40 Mounting ATEX electronics housing A0053050 9 Dimension of ATEX electronics housing. Unit of measurement mm (in) Rod sensor The rod sensor can be installed using a mounting flange and a 200 mm (7.87 in) long installation pipe (additional mounting accessories are optionally available).
Page 14
Mounting Solitrend MMP40 A0038248 10 Rod sensor mounted, rear view Rod sensor Vessel wall Mounting flange Installation pipe / extension / adapter (accessories) 6.6.1 Mounting flange ⌀55 mm (2.17 in) The mounting flange for the rod sensor can be mounted on the vessel wall.
Page 15
Solitrend MMP40 Mounting An installation pipe must first be mounted on the sensor before the mounting flange can be fixed. It is available in two different lengths and can be ordered together with the device via the "Accessory mounted" or "Accessory enclosed" section of the product order structure.
Mounting Solitrend MMP40 In addition, the sensor and ground cable can be protected with a silicone tube (not included in the delivery) A0037429 14 Example of round sensor Post-mounting check Perform the following checks after mounting the device: ...
Solitrend MMP40 Electrical connection Electrical connection Connecting requirements 7.1.1 Cable specification Connecting cables are available in different versions and lengths (depending on design). Device with 10-pin connector Connecting cables with a pre-assembled 10-pin socket on the device side are available in different standard lengths: •...
Page 18
Electrical connection Solitrend MMP40 A0037415 15 Assignment of the 10-pin connector 12 to 24 V stabilized power supply Wire color: red (RD) power supply Wire color: blue (BU) 1st analog positive (+), material moisture Wire color: green (GN) 1st analog return line (-), material moisture...
Page 19
Solitrend MMP40 Electrical connection A0044667 16 Rectangular sensor (standard version) with 10-pin cable assignment 10-pin cable with ferrules 12 to 24 V stabilized power supply Wire color: white (WH) power supply Wire color: brown (BN) 1st analog positive (+), material moisture...
Page 20
Electrical connection Solitrend MMP40 A0053676 17 ATEX electronics housing with assignment of the 10-pin cable 10-pin cable with ferrules 12 to 24 V stabilized power supply Wire color: red (RD) power supply Wire color: blue (BU) 1st analog positive (+), material moisture...
Page 21
Solitrend MMP40 Electrical connection Connection to a remote display (optional) 1 234 A0040962 18 Connection to a remote display USB (Mini B type), USB-IMP-Bridge, firmware update (only for service purposes) Socket for supply voltage and bus interface Connector for supply voltage and bus interface (included in the delivery for "remote display")
Electrical connection Solitrend MMP40 7.2.2 Connection example of 10-pin socket 6 7 8 A0037418 19 Connection example, cable with 10-pin socket (on device side) and wire end ferrules at the cable Transmitter PLC/distribution box Remote display (optional) power supply...
Solitrend MMP40 Operation options Does the supply voltage match the specifications on the nameplate? Are the connections established correctly and protected against mechanical influences? Operation options Overview of operation options RS485 IMP-Bus A0037417 20 Operation options Round sensor, short...
Page 24
Commissioning Solitrend MMP40 Commissioning Analog outputs for measured value output The measured values are output as a current signal via the analog output. The device can be set to 0 to 20 mA or 4 to 20 mA. The current output can also be set inversely to 20 to 0 mA or 20 to 4 mA for special controllers and applications.
Solitrend MMP40 Commissioning Channels of the analog outputs The analog outputs can be set differently to the following possible options: Moisture, temperature Output 1 for moisture, output 2 for material temperature. Moisture, conductivity Output 1 for moisture, output 2 for conductivity in the range from 0 to 20 mS/cm...
Page 26
Commissioning Solitrend MMP40 Operating mode The sensors are supplied from the factory with the CH mode for applications in the construction industry, and with the CA mode for general process applications. Six different operating modes are available in the C mode, depending on the application •...
Solitrend MMP40 Diagnostics and troubleshooting depending on the mineral concentration. The conductivity measuring range of the sensors here is up to 20 mS/cm, depending on the moisture content. 9.4.2 Material temperature measurement The sensor contains an integrated temperature sensor, which determines the housing temperature 3 mm below the surface of the sensor head.
Page 30
Diagnostics and troubleshooting Solitrend MMP40 A0037430 23 Example: "Concentration of material" Minimum sensor coverage 35 mm (1.38 in) In addition, in the case of non-homogeneous material flow it is possible to use the filter functions, with upper and lower limits, that are implemented in the sensor to filter out "wrong"...
Page 31
Solitrend MMP40 Diagnostics and troubleshooting If the moisture value of the sensor deviates by more than ±1 % from the laboratory value during initial commissioning, this may be due to the following: • The sensor is not correctly installed under the silo discharge hatch. The surface of the sensor must be completely covered with sand/gravel when the hatch opens.