General Specifications; Zerowire Mobile; Zerowire Mobile Accessories; Safety Features - NDS ZeroWire Mobile User Manual

Hide thumbs Also See for ZeroWire Mobile:
Table of Contents

Advertisement

General Specifications

The ZeroWire Mobile is designed to provide power to an electrical load and can be used continuously.
Note: Specifications are subject to change without notice. Contact factory for current specifications.

ZeroWire Mobile

The ZeroWire Mobile includes the following:
• ZeroWire Mobile stand with 125 mm diameter casters (5x)
• Rear-mounted fixed-position handle on the base column
• Extension column provides monitor height adjustment, and the VESA hinged mount provides monitor tilt
adjustment (see"ZeroWire Mobile Monitor Setup" on page 15)

ZeroWire Mobile Accessories

ZeroWire Mobile accessories include the following:
• ZeroWire Mobile power modules including built-in regulation (output 24 Vdc, maximum 150 W/6.25 A
continuous operation).
• ZeroWire Mobile wall-mounted Charging Station (4-bay charger; input 100–240 Vac, 4.6–2.3 A, 50/60 Hz), with
two hospital-grade AC power cords (NEMA 5-15p for North America and CEE 7/7 for EMEA).
• ZeroWire Mobile power cable for NDS 19" – 26" monitors.
• ZeroWire Mobile power cable for NDS 26" HB – 32" monitors.
• 4-bay charger service kit (part number: 90Z0176)
Please reference "ZeroWire Mobile Components" on page 13 for additional information regarding ZeroWire
Mobile system components and accessories.

Safety Features

See "Charging Station Overview" on page 17, and "Power Module Overview" on page 18.
Authentication
All ZeroWire Mobile power modules are factory programmed with a password code used for verification that the
power module is a genuine NDS product each time the unit is docked, preventing use of counterfeit devices.
ZeroWire Mobile is a medical device, therefore, power module authentication is a critical safety technology to
protect both the user and the patient.
Dock Detection
External contacts of power modules are enabled only when docked. When power modules are undocked, the
dock detection feature isolates internal cell voltage from the external contacts for protection against electric
shock and short circuit hazards.
Temperature Monitoring
The internal temperature of docked power modules is continuously monitored. If the internal temperature
exceeds the safe operating temperature range, charging or discharging is suspended until the temperature is
restored to a safe level. See "Operation" on page 7.
Overcharge Prevention
Charging is automatically terminated when power modules are fully charged.