Electric Charge For Detecting Discharge Inception Voltage (Inception Voltage) - Kikusui TOS9301PD User Manual

Electrical safety analyzer
Table of Contents

Advertisement

Partial Discharge Test | Set test conditions
Electric charge for detecting discharge inception voltage
(Inception Voltage)
This function detects the voltage when the specified electric charge is measured the first time after starting
a test as discharge inception voltage.
Discharge inception
voltage [V]
Inception Voltage [pC]
1
On the HOME Menu screen, press Coulomb repeatedly until Coulomb is selected.
Each press of the key switches between Coulomb and Measure.
2
Press Inception Voltage, and the use the numeric keypad or the rotary knob to enter
a value.
The setting range varies according to the value set to Range
Setting range (100 pC range): 10.0 pC to 100.0 pC
Setting range (1000 pC range): 100.0 pC to 1 000.0 pC
Setting range (10000 pC range): 1000 pC to 10 000 pC
This completes the setting.
118
User's Manual
V, C
0 V, 0 pC
Applied voltage
Electric charge
.
(p.117)