Terminal Assignment - ABB CoriolisMaster FCB400 Series Manual

Hide thumbs Also See for CoriolisMaster FCB400 Series:
Table of Contents

Advertisement

CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H

Terminal assignment

A Transmitter
Figure 28: Electrical connection
Connections for the power supply
AC voltage
Terminal
Function / comments
L
Phase
N
Neutral conductor
PE /
Protective earth (PE)
Potential equalization
DC voltage
Terminal
Function / comments
1+
+
2-
-
PE /
Protective earth (PE)
Potential equalization
B Sensor
Connections for inputs and outputs
Terminal
Function / comments
Uco / 32
Current output 4 to 20 mA- / HART® output, active
or
31 / 32
Current output 4 to 20 mA- / HART® output, passive
41 / 42
Passive digital output DO1
51 / 52
Passive digital output DO2
V1 / V2
Plug-in card, slot OC1
V3 / V4
Plug-in card, slot OC2
For details, see Optional plug-in cards on page 40.
Connecting the signal cable
Only for remote mount design.
The sensor housing and transmitter housing must be connected
to potential equalization.
Terminal
Function / comments
U
Sensor power supply
FE
GND
Ground
A
Data line
B
Data line
Functional earth / Shielding