CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H
A Signal cable to sensor
1 Terminals for power supply
2 Terminals for inputs and outputs (base device)
3 Terminals for signal cable
4 Terminals for inputs and outputs (plug-in cards)
Figure 52:
Electrical connection to transmitter in remote mount design [example, dimensions in mm (in)]
NOTICE
If the O-ring gasket is seated incorrectly or damaged, this
may have an adverse effect on the housing protection class.
Follow the instructions in Opening and closing the housing
on page 37 to open and close the housing safely.
Terminal
ABB signal cable
3KQZ407123U0100
GND
Blue
U
White
FE
A
Yellow
B
Orange
Single-compartment housing
5 Terminal for potential equalization
6 LCD indicator
7 Bracket for LCD indicator (park position)
8 Terminal for protective earth / cable shields
Observe the following points when connecting to an electrical
supply:
HELKAMA signal cable
20522
Blue (4)
white (3)
Blue (2)
white (1)
Lead the cable for the power supply and the signal inputs
and outputs into the housing as shown.
The signal cable to the sensor is connected in the lower
connection area of the transmitter.
Connect the cables in accordance with the electrical
connection diagram. If present, connect the cable shielding
to the earthing clamp provided.
Use wire end ferrules when connecting.
After connecting the power supply, terminal cover
must be installed.
Close unused cable entries using suitable plugs.
59