CoriolisMaster FCB400, FCH400 CORIOLIS MASS FLOWMETER | CI/FCB400/FCH400-EN REV. H
Wiring with different network topologies
A Daisy-chain
B Ring
Figure 57: Connection topologies
Ethernet Option Cards are designed only for use in hazardous applications Zone 2 / Division 2 or general purpose areas.
The output circuits are designed so that different topologies such as daisy chain or point to point can be connected.
See Installation diagram for detailed information.
It is not permitted to combine both topologies.
Ethernet communication is only available for installations in Zone 2/Division 2 or general purpose
The rated voltage of these non-intrinsically safe circuits are UM = 57 V.
Topology
Star
C Star
No. Ethernet cables
No. wires in
connected
Ethernet cable
1
1
1
1
PoE
Port
Clamp
Function
4
No
1
1
1
3
4
8
No
1
1
2
3
4
2
1
2
3
4
4
Yes
1
1
2
3
4
8
Yes
1
1
2
3
4
2
1
2
3
4
65
Cable
RD+
white / orange
RD−
orange
TD+
white / green
TD−
green
RD+
white / orange
RD−
orange
TD+
white / green
TD−
green
Spare 1+
white / blue
Spare 1−
blue
Spare 2+
white / brown
Spare 2−
brown
Recommendation:
Use cable with 8 wires
RD+
white / orange
RD−
orange
TD+
white / green
TD−
green
Spare 1+
white / blue
Spare 1−
blue
Spare 2+
white / brown
Spare 2−
brown

Hide quick links: