7.4 OFF BLOCK ZERO (AUTOMATIC PROBE ZERO)
1. When the ZERO PROBE screen is displayed, remove all couplant from the face of the
transducer and check that the wearface of the transducer is clean and free of any debris.
2. Press OK (or ESC to cancel).
When the automatic probe zero is completed the measurement screen is displayed.
Your gauge should now be calibrated.
7.5 ON BLOCK ZERO (MANUAL PROBE ZERO)
When this zero probe option is selected a warning screen is displayed informing you that you must
zero the probe before taking measurements.
1. Press MEAS to display the measurement screen.
2. Remove all couplant from the face of the transducer and check that the wearface is clean and
free of any debris.
3. Apply a drop of couplant on the transducer and place the
transducer as shown in diagram.
Note: Disregard the value displayed; it is not important. What is
important is accurately performing these steps to ensure
reliability of the zero calculation.
4. Press MEAS, scroll to the PRB menu, scroll to the ZERO
PROBE function and then press ENTER.
The ZERO PROBE screen is displayed.
5. UP and DOWN to toggle coating on/off.
6. Press OK (or ESC to cancel).
When the manual probe zero is completed the measurement
screen is displayed.
7. Remove the transducer from the probe zero disk.
Your gauge should now be calibrated.
7.6 TRANSDUCER - ZEROING FOR COATING ONLY
MEASUREMENTS
Your gauge features an additional zero coating feature for use when measuring coatings to a high
accuracy is required
Note: For this procedure, a coating transducer must be fitted to the gauge.