Mains Supply

The following table contains an overview on the electrical contact data of the Sorvall RC-4. This data is
to be taken into consideration when selecting the mains connection socket.
Table 1-4. Electrical Connection Data
Cat.
75004473
75004477
75004481
75004481
75004481
Thermo Scientific
1...9
1 = slow, ... 9 = fast Acceleration / Deceleration Curve
Adjustable from 300 rpm to10,000 rpm in 10 rpm increments
After actuating the RCF Switch Key, the RCF Value can be entered.
Adjustable in Seconds up to 9 min 59 sec,
in the Range of 10 min up to 99 min
in Minutes, "hold"-Mode: Continuous Operation
In Seconds Increments
Adjustable from -9°C to +40°C, in 1°C Increments.
The Speed Display will read End.
Automatic unlocking via Key
(Unlocking in Case of Power Failure:
see
"Mechanical Emergency Door Release"
Key
Key
Pressing the
permissible Speed of Rotor; upon Key release Centrifuge stops with maximum
Deceleration Power.
Alternating Display
max. Speed of the Rotor)
Door has not been lifted off the Lock during Opening: Display
Lifting of Door required)
General Centrifuge Malfunction (Error Messages with ERROR Codes,
see"Troubleshooting by User"
Mains
Frequency Rated
Voltage
230 V
50 Hz
120 V
60 Hz
200 V
50 / 60 Hz
208 V
60 Hz
230 V
60 Hz
Feature
Free Program Selection
Operating Memory
Memory Capacity for set Values
PULSE-Mode
Key activates maximum Acceleration up to the maximum
rotor
/maximum Speed or RCF (Speed selected exceeds
on
page
Power
Consumption
Current
12 A
2.7 kW
16 A
1.8 kW
14.5 / 16 A
2.7 kW
14.5 A
2.7 kW
14.5 A
2.7 kW
1
Introduction and Description
on
page
6-2)
Lift Lid
6-3)
Equipment fuse
Building fuse
16 AT
16 AT
16 AT
20 AT
16 AT
20 AT
16 AT
20 AT
16 AT
20 AT