Thermo Scientific
Diagnostik Rotor
Instruction Manual
50135351-b • 07 / 2020
Visit us online to register your warranty:
thermofisher.com/labwarranty

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Cryofuge 5500i and is the answer not in the manual?

Questions and answers

Summary of Contents for Thermo Scientific Cryofuge 5500i

  • Page 1 Thermo Scientific Diagnostik Rotor Instruction Manual 50135351-b • 07 / 2020 Visit us online to register your warranty: thermofisher.com/labwarranty...
  • Page 2 WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:...
  • Page 3: Table Of Contents

    Improper Loading ....................4-3 Chapter 5 Maintenance and Care ......................5-1 Cleaning Intervals ....................5-2 Cleaning ........................5-2 Disinfection ......................5-3 Decontamination ..................... 5-4 Service of Thermo Fisher Scientific ................5-5 RCF-Values .......................A-1 Chemical Compatibility Chart ................B-1 Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 4 Contents 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 5: Preface

    WARNING means that injuries or material damage or contamination could occur. CAUTION means that material damage could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 6: Precautions

    Please pay particular attention to the following aspects: • Rotor installation: Check that the rotor is locked properly into place before operating the centrifuge. • Always balance the samples. Maximum sample density at maximum speed: 1.2 ---------- - 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 7: Rotor Data

    Rotor Data Contents • “Technical Data” page 1-2 Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 8: Technical Data

    Rotor Data Technical Data Technical Data Table 1-1. 230 V, 50 / 60 Hz Centrifuge Cryofuge 5500i Cryofuge 5500i Multifuge 4KR Catalog Number 75004471 75004479 75004461 Weight of empty Rotor [kg] 10.7 10.7 10.7 Maximum permissible Load [ g ]...
  • Page 9 Rotor Data Technical Data Table 1-3. 120 V, 60 Hz Centrifuge Cryofuge 5500i Catalog Number 75004475 Weight of empty Rotor [kg] 10.7 Maximum permissible Load [ g ] 2 x 2000 Maximum Speed n [ rpm ] 3500 Maximum RCF-Value at n 2547 Radius max.
  • Page 10 Acceleration / Braking Time [ s ] 65 / 65 Sample Heating at n [° C] referred to Ambient < 0 Temperature of 23-25 °C, Run Time 60 Minutes Max Speed n [ rpm ] at 4 °C 3500 Aerosol-tight Autoclavable 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 11: Accessories

    Accessories Contents • “Accessory“ page 2-2 Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 12 [mm] Tubes and Rack [mm] Rotor Hitachi 118 x 20 2 x 10 75006416 Olympus 177 x 20 2 x 6 75006417 Sysmex 200 x 25 2 x 6 75006418 Dade-Behring 2 x 2 75006419 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 13: Rotor Installation

    Rotor Installation Contents • “Rotor Installation” page 3-2 “Removing the Rotor” page 3-2 • Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 14: Removing The Rotor

    3. At the same time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle. Make sure not to tilt the rotor while doing this operation. 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 15 Rotor Loading Contents • “Before a Run” page 4-2 “Proper Loading” page 4-2 • • “Improper Loading” page 4-3 Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 16: Chapter 4 Rotor Loading

    To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times. The buckets and micro plate carriers must always be fully loaded with racks. If there are racks missing, the running racks might fall over and cause a rotor failure due to imbalance. 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 17: Improper Loading

    Rotor Loading Improper Loading Improper Loading Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 18 Rotor Loading Improper Loading 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 19: Maintenance And Care

    Maintenance and Care Contents • “Cleaning Intervals” page 5-2 “Cleaning” page 5-2 • • “Disinfection” page 5-3 “Decontamination” page 5-4 • • “Autoklavieren” Seite 6-5 “Service of Thermo Fisher Scientific” page 5-5 • Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 20: Cleaning Intervals

    • After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (70009824). Also treat the cavities with oil. • Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing down. 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 21: Disinfection

    Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not kept at risk. Decontaminate the affected parts immediately. Take other precautions if needed. Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 22: Decontamination

    Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be. 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 23: Service Of Thermo Fisher Scientific

    Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are performed for free during the warranty period and afterwards for a charge. This is only valid if the centrifuge has only been maintained by a Thermo Fisher Scientific service technician. Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 24 Maintenance and Care Service of Thermo Fisher Scientific 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 25: Rcf-Values

    14.1 1100 2400 18.6 14.1 1198 2500 18.6 14.1 1300 2600 18.6 14.1 1406 1066 2700 18.6 14.1 1516 1149 2800 18.6 14.1 1630 1236 2900 18.6 14.1 1749 1326 3000 18.6 14.1 1872 1419 Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 26 RCF-Values Speed rpm RCF R RCF R 3100 18.6 14.1 1998 1515 3200 18.6 14.1 2129 1614 3300 18.6 14.1 2265 1717 3400 18.6 14.1 2404 1822 3500 18.6 14.1 2547 1931 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 27: Chemical Compatibility Chart

    U S S M S S U S S S S S S S S S U S S S S S S S S S S Cesium Acetate S S S - S S S S - S S - S S S S S S M S S S Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 28 U U U U S U U U S S M M S U U U - Hydrofluoric Acid (50 %) U U U U - U U M - U M U U M U U U - U S U U U - Hydrochloric Acid (conc.) 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 29 S S S S S S S M S M S S - Lactic Acid (20 %) S U - S M - U S M S S S S M M S M - N-Butyl Alcohol Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 30 S S S U S S S Sulfosalicylic Acid U S U S S U U - S U S U - S S S S S S S S S M S S S S Nitric Acid (10 %) 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 31 S S S S S S S S S S S S S S S S S S S S S S S Citric Acid (10 %) M S S M S S M S S S S S S S S S M S S S S S S S S S S Polyethyleneterephthalate Thermo Scientific 75006480 Diagnostik Rotor...
  • Page 32 Performance unknown; suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. No organized chemical resistance data exists for materials under the stress of centrifugation. When in doubt we recommend pretesting sample lots. 75006480 Diagnostik Rotor Thermo Scientific...
  • Page 34 Thermo Electron LED GmbH Zweigniederlassung Osterode Am Kalkberg, 37520 Osterode am Harz Germany thermofisher.com/rotors © 2012-2020 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries unless otherwise indicated. Delrin, TEFLON, and Viton are registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a registered trademark of Hongye CO., Ltd.