Summary of Contents for Thermo Scientific Cryofuge 5500i
Page 1
Thermo Scientific Diagnostik Rotor Instruction Manual 50135351-b • 07 / 2020 Visit us online to register your warranty: thermofisher.com/labwarranty...
Page 2
WEEE Conformity This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2012/19/EU. It is marked by the following symbol:...
WARNING means that injuries or material damage or contamination could occur. CAUTION means that material damage could occur. This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe. Thermo Scientific 75006480 Diagnostik Rotor...
Please pay particular attention to the following aspects: • Rotor installation: Check that the rotor is locked properly into place before operating the centrifuge. • Always balance the samples. Maximum sample density at maximum speed: 1.2 ---------- - 75006480 Diagnostik Rotor Thermo Scientific...
Rotor Data Technical Data Technical Data Table 1-1. 230 V, 50 / 60 Hz Centrifuge Cryofuge 5500i Cryofuge 5500i Multifuge 4KR Catalog Number 75004471 75004479 75004461 Weight of empty Rotor [kg] 10.7 10.7 10.7 Maximum permissible Load [ g ]...
Page 9
Rotor Data Technical Data Table 1-3. 120 V, 60 Hz Centrifuge Cryofuge 5500i Catalog Number 75004475 Weight of empty Rotor [kg] 10.7 Maximum permissible Load [ g ] 2 x 2000 Maximum Speed n [ rpm ] 3500 Maximum RCF-Value at n 2547 Radius max.
Page 10
Acceleration / Braking Time [ s ] 65 / 65 Sample Heating at n [° C] referred to Ambient < 0 Temperature of 23-25 °C, Run Time 60 Minutes Max Speed n [ rpm ] at 4 °C 3500 Aerosol-tight Autoclavable 75006480 Diagnostik Rotor Thermo Scientific...
3. At the same time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle. Make sure not to tilt the rotor while doing this operation. 75006480 Diagnostik Rotor Thermo Scientific...
To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times. The buckets and micro plate carriers must always be fully loaded with racks. If there are racks missing, the running racks might fall over and cause a rotor failure due to imbalance. 75006480 Diagnostik Rotor Thermo Scientific...
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (70009824). Also treat the cavities with oil. • Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing down. 75006480 Diagnostik Rotor Thermo Scientific...
Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not kept at risk. Decontaminate the affected parts immediately. Take other precautions if needed. Thermo Scientific 75006480 Diagnostik Rotor...
Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be. 75006480 Diagnostik Rotor Thermo Scientific...
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are performed for free during the warranty period and afterwards for a charge. This is only valid if the centrifuge has only been maintained by a Thermo Fisher Scientific service technician. Thermo Scientific 75006480 Diagnostik Rotor...
Page 24
Maintenance and Care Service of Thermo Fisher Scientific 75006480 Diagnostik Rotor Thermo Scientific...
U S S M S S U S S S S S S S S S U S S S S S S S S S S Cesium Acetate S S S - S S S S - S S - S S S S S S M S S S Thermo Scientific 75006480 Diagnostik Rotor...
Page 28
U U U U S U U U S S M M S U U U - Hydrofluoric Acid (50 %) U U U U - U U M - U M U U M U U U - U S U U U - Hydrochloric Acid (conc.) 75006480 Diagnostik Rotor Thermo Scientific...
Page 29
S S S S S S S M S M S S - Lactic Acid (20 %) S U - S M - U S M S S S S M M S M - N-Butyl Alcohol Thermo Scientific 75006480 Diagnostik Rotor...
Page 30
S S S U S S S Sulfosalicylic Acid U S U S S U U - S U S U - S S S S S S S S S M S S S S Nitric Acid (10 %) 75006480 Diagnostik Rotor Thermo Scientific...
Page 31
S S S S S S S S S S S S S S S S S S S S S S S Citric Acid (10 %) M S S M S S M S S S S S S S S S M S S S S S S S S S S Polyethyleneterephthalate Thermo Scientific 75006480 Diagnostik Rotor...
Page 32
Performance unknown; suggest testing, using sample to avoid loss of valuable material. Chemical resistance data is included only as a guide to product use. No organized chemical resistance data exists for materials under the stress of centrifugation. When in doubt we recommend pretesting sample lots. 75006480 Diagnostik Rotor Thermo Scientific...