Temperature Adaptation During Standstill; Short-Term Centrifugation - Thermo Scientific Sorvall RC-4 Instruction Manual

Hide thumbs Also See for Sorvall RC-4:
Table of Contents

Advertisement

Stopping the Centrifugation Program
With preset Run Time
Usually the run time is preset and you only have to wait until the centrifuge stops automatically when
the preset time limit expires.
As soon as the speed drops to zero, the message
you can open the door and remove the centrifuge material.
You can also stop the centrifuging program manually at any time by pressing the
Continuous Operation
If you selected continuous operation (see
the centrifuge manually. Press the
the designated rate. As soon as the speed drops to zero, the message
pressing the

Temperature Adaptation during Standstill

The temperature cannot be adapted until the rotor has been positively identified. For this, the
centrifuge must have accelerated to over 300 rpm. The speed display will then show End.
When the rotor is not recognized (door closed and
centrifuge responds by ensuring that the sample cannot freeze regardless of the rotor being used.

Short-term Centrifugation

For short-term centrifuging, the Sorvall RC-4 has a PULSE- function.
By holding down the
The centrifuge accelerates and brakes at maximum power. Any rpm or RCF entered beforehand is
overridden.
Note The centrifuge accelerates to maximum speed, regardless of which rotor was installed.
Check carefully whether you have to maintain a certain speed for your application.
During the acceleration process, time is counted forwards in seconds. The reading stays displayed until
the centrifuge door is opened.
Thermo Scientific
key on the control panel. The centrifuge will be decelerated at
key, you can open the door and remove the sample.
key, spinning will start and continue until the key is let go.
End
will appear in the display. By pressing the
"Continuous Operation"
key not yet pressed, speed display 0), the
Temperature Adaptation during Standstill
key.
on
page
4-6), you will have to stop
End
will appear in the display. By
Sorvall RC-4
4
Operation
key,
4-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents