A.6 First Alignment of the Light Lever
This procedure assumes that the light lever was cor-
rectly aligned at the factory or by an authorized agent.
1.
Place a sample in the sample holder, and
slide the sample holder under the light lever.
2.
Place the probe holder in the light lever sen-
sor with a contact or vibrating mode probe
installed.
3.
Adjust the optical microscope until the probe/
substrate can be viewed on screen.
4.
Center the laser and detector translation
mechanisms with thumb screws.
5.
Turn on the laser.
6.
Use a flathead screwdriver to rotate the mirror
until the laser spot is on the back side of the
cantilever.
V 1.1 / LS-AFM Users Guide
Translation thumb screws
Set screw holding laser
Rotating mirror mechanism
Appendix A-5