3.2.2.3 Scan Setup

All of the parameters required for scanning are presented in the Scan Setup window. Once
scanning is started, these parameters may not be changed. The parameters are:
1.
Type: The scan can be either 2-D, which scans just the x-axis, or 3-D, which scans both
the x- and y-axis.
2.
Scan Rate: The rate in Hertz that the
probe is scanned across the surface
(max 2 Hz).
3.
Scan Lines: The number of lines in the
y-axis and the number of pixels mea-
sured in the x-axis.
4.
Scan Size: The range of the scan,
from 0.1 micron to either 15 or 50
microns (depending on scanner size).
5.
Samples/Pixel: The number of data
points signal averaged per pixel while
a scan is being taken.
6.
Rotation: A scan can be completed
with 0° or 90° rotation.
7.
Center: The center of the scan in x and y coordinates.
8.
Current Y/X Position: The position of the probe along the y-axis as a scan is taken at
0°, along the x-axis for a 90° scan.
9.
Scan Complete: This indicator turns green when a scan is completed.
10.
Zoom Box: This shows the position and size of the scan in a grid. The scan area must
be within the zoom box area.

3.2.2.4 Z Feedback

All of the parameters for maintaining feedback are provided
here. The parameters are:
1.
Gain: Scale factor for the feedback control signal.
2.
Proportional: Scale factor for the proportional term in the
PID feedback controller.
3.
Integral: Scale factor for the integral term in the PID
feedback controller.
Section 3-8
V 1.1 / LS-AFM Users Guide