Associated Values for Specific Registry Entries

Associated Values for Specific Registry Entries
Device Status
Entries
Device State
Entries (LSByte)
Device SubState
Entries (MSByte)
Data Packet Status
Entries
Airnet® II 4-Channel Particle Sensors Operations Manual

Table B-5 Associated values for specific registry entries

0x0001
Data Collection
0x0002
Data Available
0x0004
Data Clear
0x0008
Reset
0x0010
Green Status
0x0020
Red Status
0x0040
External Alarm
0x0080
IEEE Float mode
0
Idle
1
Sampling
2
Maintenance
3
Error
0
Idle
0
Sampling
1
Sampling
0
Maintenance
1
Maintenance
0
Error
0x0001
Laser Good
0x0002
Flow Good
0x0004
Hardware Good
Appendix B : Modbus Protocol
Matches the common coil settings
n/a
n/a
Tare
n/a
Calibration
n/a