Chapter 6
Troubleshooting
Use this chapter to help troubleshoot issues with the Airnet II 4-Channel particle
sensor, such as power issues, communication messages, errors, leaks, and particle and
transient-induced counts.
Troubleshooting Matrix
Problem
Symptoms
No power
No lights
or
No communications
Communications
Unable to view data
not established –
from working sensor
Facility Net
displays
"Communications
Error".
Laser status error
Bad laser status
indication
Airnet® II 4-Channel Particle Sensors Operations Manual
Possible Cause
Try...
Power supply not
Check the seating of the power supply
connected to a power
cable in the unit.
source
Verify that the power supply is plugged
in.
Electrical power
Check the electrical outlet for power,
source fault
including any power strips or surge
protectors.
Power supply faulty
Try a different power supply.
If using PoE
Verify power is being supplied
Ethernet cable
Replace the Ethernet cable that connects
problem
the sensor to an Ethernet hub.
Unit not configured
Re-configure the unit. First check the
correctly
sensor's IP Address and the computer's
IP Address.
Computer hardware
Substitute a computer with known good
or software problem
Ethernet capability.
Refer to your computer's Ethernet card
documentation to correct the fault.
Laser dead or
Check status response for detector
detector failure
failure messages.
6-1