Chapter 6: Replacing Assemblies
To set the calibration factors after replacing the acquisition board
Parity: None: 8 bits
Language: SCPI
c Pull down the Utilities menu and select Calibration.
d Uncheck to Cal Memory Protect box to allow calibration.
e Click on Start to start the calibration procedure.
Follow the on-screen instructions as calibration proceeds.
122