Chapter 6: Replacing Assemblies
To remove and replace the acquisition board assembly
Carefully lift the back of the board to separate the acquisition board connectors from
7
the power distribution board.
Lift the board back from the front panel until the BNC connectors clear the front panel,
8
then lift the board away from the chassis.
Figure 6-20
Connectors
Separate acquisition board from power distribution board
To re-install the acquisition board assembly, reverse the above procedure.
9
Now the calibration factors must be set. See "To set the calibration factors after
10
replacing the acquisition board" on page 121.