Chapter 6: Replacing Assemblies
To remove and replace the AutoProbe assembly
d Disconnect the mylar flex cable from the AutoProbe board.
Figure 6-6
Disconnect mylar
flex cable here
Pushing out the AutoProbe faceplate
To replace the AutoProbe assembly, reverse the above procedure.
3
108
See See "To disconnect and connect Mylar flex cables" on page 106.