At rest, the display reads: 
Press ENTER. 
Step 1: 
Display will read: 
Scan the barcode on the probe electrical 
connector and place into the wall‐mounted 
Step 2: 
Connector Holder.    
Display will read: 
Insert probe into disinfection chamber and 
align blue tape on probe cable with Probe 
Alignment Mark between two Probe Cable 
Holders (as shown below). Slide the probe 
Step 3:  
cable into the Probe Cable Clip. Place probe 
cable into cable holder. 
Display will read: 
Entering Exam ID is optional, but you must 
scan ID or press ENTER to continue. 
Step 4: 
Display will read: 
 
 
 
®
ASTRA
 VR (50230_E) 
 
ASTRA
           QUICK START INSTRUCTIONS 
 
 
 
 
 
 
Alignment of Probe Cable
 
®
 VR 
Ready for Use 
Press Enter to Start 
Scan Probe 
Insert Probe 
Scan Exam ID 
ENTER = skip 
Close Door or 
Scan Second Probe 
 
 

Advertisement

loading
Need help?

Need help?

Do you have a question about the ASTRA VR and is the answer not in the manual?

Questions and answers

Summary of Contents for Civco ASTRA VR

  • Page 1 ® ASTRA  VR             QUICK START INSTRUCTIONS      Ready for Use    At rest, the display reads:    Press Enter to Start      Press ENTER.  Step 1:  Scan Probe  Display will read:  Scan the barcode on the probe electrical    connector and place into the wall‐mounted  Step 2:  Insert Probe  Connector Holder.     Display will read:  Insert probe into disinfection chamber and  align blue tape on probe cable with Probe  Alignment Mark between two Probe Cable    Scan Exam ID  Holders (as shown below). Slide the probe  Step 3:   ENTER = skip  cable into the Probe Cable Clip. Place probe  cable into cable holder.  Display will read:  Entering Exam ID is optional, but you must  scan ID or press ENTER to continue.  Close Door or  Step 4:  Scan Second Probe  Display will read: ...
  • Page 2 ® ASTRA  VR  INSTRUCTIONS DE DÉMARRAGE      Ready For Use     (Prêt à l’utilisation)    À l’état de veille, l’écran affiche :  Press ENTER to Start (Appuyer sur    ENTRÉE pour commencer)  Appuyer sur ENTER (ENTRÉE).  Étape 1:    Scan Probe   L’écran affichera :  (Scanner la sonde)  Scanner le code à barres sur le connecteur  électrique de la sonde et le placer dans le  Étape 2:    Insert Probe   support mural du connecteur.  (Insérer la sonde)  L’écran affichera :  Insérer la sonde dans la chambre de  désinfection et aligner le ruban adhésif bleu  sur le câble de la sonde avec la marque  Scan Exam ID   d’alignement de la sonde, entre deux  (Scanner l’ID d’examen)    Étape 3:   ENTER = skip (ENTRÉE = Ignorer)  supports du câble (tel qu’illustré ci‐dessous).  Faire glisser le câble de la sonde dans la pince  à câble. Placer le câble dans son support.  L’écran affichera : ...
  • Page 3 Install second probe (if required) following same    instructions or close door. Barcode scanner will  Scan Operator ID  remain illuminated until door is closed or a  Step 5:    ENTER = Skip  second probe is scanned.      Display will read:  Scanning Operator ID or pressing ENTER to Skip    Filling chamber with   will start the disinfection cycle.    Step 6:  HLD, please wait….  Display will read:    Once level is reached, open disinfection    chamber door and dip an MRC test strip into  Test MRC  Step 7:  HLD, holding it submerged for 2 seconds.  PASS <         > FAIL  Remove test strip and close door.  Display will read:  Compare test strip to color chart on  manufacturer’s bottle.    Test MRC  Step 8:  PASS <         > FAIL  Display will read:             Press PASS or FAIL using corresponding arrows      buttons.  If PASS is pressed, display will read:  Disinfecting  Step 9: ...
  • Page 4 Installer la seconde sonde (si nécessaire) en    suivant les mêmes instructions ou fermer la  Scan Operator ID     porte. Le scanneur de code à barres demeurera    (Scanner l’ID d’utilisateur) Étape 5:    allumé jusqu’à ce que la porte soit fermée ou  ENTER = skip   qu’une seconde sonde soit scannée.     (ENTRÉE = Ignorer) L’écran affichera :    Le fait de scanner l’ID d’utilisateur ou d’appuyer  Filling chamber with (Remplissage  sur ENTER (ENTRÉE) pour passer à l’étape    de la chambre avec)  Étape 6:  suivante déclenchera le cycle de désinfection.   HLD, please wait…. (du  L’écran affichera :    désinfectant, veuillez attendre...)   Une fois le niveau atteint, ouvrir la porte de la    chambre de désinfection et tremper pendant    Test MRC (Test CMR) deux secondes la bandelette de test CMR dans  Étape 7:  PASS <         > FAIL   le désinfectant. Retirer la bandelette de test et    (SUCCÈS <          > ÉCHEC) fermer la porte. ...