Page 1
® ASTRA VR QUICK START INSTRUCTIONS Ready for Use At rest, the display reads: Press Enter to Start Press ENTER. Step 1: Scan Probe Display will read: Scan the barcode on the probe electrical connector and place into the wall‐mounted Step 2: Insert Probe Connector Holder. Display will read: Insert probe into disinfection chamber and align blue tape on probe cable with Probe Alignment Mark between two Probe Cable Scan Exam ID Holders (as shown below). Slide the probe Step 3: ENTER = skip cable into the Probe Cable Clip. Place probe cable into cable holder. Display will read: Entering Exam ID is optional, but you must scan ID or press ENTER to continue. Close Door or Step 4: Scan Second Probe Display will read: ...
Page 2
® ASTRA VR INSTRUCTIONS DE DÉMARRAGE Ready For Use (Prêt à l’utilisation) À l’état de veille, l’écran affiche : Press ENTER to Start (Appuyer sur ENTRÉE pour commencer) Appuyer sur ENTER (ENTRÉE). Étape 1: Scan Probe L’écran affichera : (Scanner la sonde) Scanner le code à barres sur le connecteur électrique de la sonde et le placer dans le Étape 2: Insert Probe support mural du connecteur. (Insérer la sonde) L’écran affichera : Insérer la sonde dans la chambre de désinfection et aligner le ruban adhésif bleu sur le câble de la sonde avec la marque Scan Exam ID d’alignement de la sonde, entre deux (Scanner l’ID d’examen) Étape 3: ENTER = skip (ENTRÉE = Ignorer) supports du câble (tel qu’illustré ci‐dessous). Faire glisser le câble de la sonde dans la pince à câble. Placer le câble dans son support. L’écran affichera : ...
Page 3
Install second probe (if required) following same instructions or close door. Barcode scanner will Scan Operator ID remain illuminated until door is closed or a Step 5: ENTER = Skip second probe is scanned. Display will read: Scanning Operator ID or pressing ENTER to Skip Filling chamber with will start the disinfection cycle. Step 6: HLD, please wait…. Display will read: Once level is reached, open disinfection chamber door and dip an MRC test strip into Test MRC Step 7: HLD, holding it submerged for 2 seconds. PASS < > FAIL Remove test strip and close door. Display will read: Compare test strip to color chart on manufacturer’s bottle. Test MRC Step 8: PASS < > FAIL Display will read: Press PASS or FAIL using corresponding arrows buttons. If PASS is pressed, display will read: Disinfecting Step 9: ...
Page 4
Installer la seconde sonde (si nécessaire) en suivant les mêmes instructions ou fermer la Scan Operator ID porte. Le scanneur de code à barres demeurera (Scanner l’ID d’utilisateur) Étape 5: allumé jusqu’à ce que la porte soit fermée ou ENTER = skip qu’une seconde sonde soit scannée. (ENTRÉE = Ignorer) L’écran affichera : Le fait de scanner l’ID d’utilisateur ou d’appuyer Filling chamber with (Remplissage sur ENTER (ENTRÉE) pour passer à l’étape de la chambre avec) Étape 6: suivante déclenchera le cycle de désinfection. HLD, please wait…. (du L’écran affichera : désinfectant, veuillez attendre...) Une fois le niveau atteint, ouvrir la porte de la chambre de désinfection et tremper pendant Test MRC (Test CMR) deux secondes la bandelette de test CMR dans Étape 7: PASS < > FAIL le désinfectant. Retirer la bandelette de test et (SUCCÈS < > ÉCHEC) fermer la porte. ...
Need help?
Do you have a question about the ASTRA VR and is the answer not in the manual?
Questions and answers