Page 1
CARE™ R EPLACEMENT ERTICAL DJUSTMENT EFERENCE UIDE FOR USE WITH: LASSIC TEPPER EX3™ S TEPPER ™ S ULTI URPOSE ORKSTATION TEPPER...
Page 3
AccuCARE™ Replacement Kit English REPLACEMENT KIT - VERTICAL ADJUSTMENT KNOB NOTE: Replacement parts are only for use with CIVCO Classic Stepper, EX3™ Stepper, and Multi-Purpose Workstation™ Stepper. Please see your equipment user guide for additional warnings and full instructions for use.
Page 4
English AccuCARE™ Replacement Kit REPLACING VERTICAL ADJUSTMENT KNOB Invert stepper so vertical adjustment knob is facing upward and set screw within stepper crossbar is visible. Using 5/16-inch hex wrench, remove set screw from crossbar. Remove existing vertical adjustment knob by turning counter-clockwise until knob releases from stepper.
Page 5
AccuCARE™ Replacement Kit English Insert new knob into place and turn until secure. Apply Loctite Blue 242 adhesive to threads of set screw. Align set screw with groove and insert into crossbar. Tighten until snug, then loosen just enough to allow free rotation of vertical adjustment knob.