At rest, the display reads:
Press ENTER.
Step 1:
Display will read:
Scan barcode on probe electrical connector
and place into the wall-mounted Connector
Step 2:
Holder.
Display will read:
Open door and insert pre-cleaned probe into
disinfection chamber, placing probe handle
into Probe Handle Holder. Loop probe cable
Step 3:
over cable strain relief.
Display will read:
Entering Exam ID is optional, but you must
scan ID or press ENTER to continue.
Step 4:
Display will read:
Install second probe (if required) following
same instructions or close door. Barcode
Step 5:
scanner will remain illuminated until door is
closed or a second probe is scanned.
Display will read:
ASTRA TEE (50229_F)
ASTRA TEE
QUICK START INSTRUCTIONS
Ready for Use
Press Enter to Start
Scan Second Probe
Scan Operator ID
ENTER = Skip
®
Scan Probe
Insert Probe
Scan Exam ID
ENTER = skip
Close Door or

Advertisement

loading
Need help?

Need help?

Do you have a question about the ASTRA TEE and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Civco ASTRA TEE

  • Page 1 Install second probe (if required) following same instructions or close door. Barcode Scan Operator ID Step 5: scanner will remain illuminated until door is ENTER = Skip closed or a second probe is scanned. Display will read: ASTRA TEE (50229_F)
  • Page 2 L’écran affichera: La saisie de l'identifiant d'examen est facultative, mais vous devez scanner Close Door or (Fermer la l'identifiant ou appuyer sur ENTER porte ou) Étape 4: (ENTRÉE) pour continuer. Scan Second Probe L’écran affichera: (Scanner deuxième sonde) ASTRA TEE (50229_F)
  • Page 3 Probe to test first? Left <- -> Right probe to test first using the corresponding arrows. Conduct leak testing according to manufacturer’s instructions. Press PASS or FAIL using the corresponding arrows. Repeat process with other probe. Data will be logged. ASTRA TEE (50229_F)
  • Page 4 Display will read: Once the countdown timer is Test MRC (Test CMR) complete (or skipped), the results of Étape 10: PASS < > FAIL (SUCCÈS < the MRC test may be read and > ÉCHEC) documented. Display will read: ASTRA TEE (50229_F)
  • Page 5 On successful completion of disinfection and Successful Cycle Step 14: rinse cycles, display will read: Remove Probe(s) The probe(s) are disinfected and should be Ready for Use removed. Step 15: Press Enter to Start After removal, display will read: ASTRA TEE (50229_F)
  • Page 6 Remove Probe(s) (Retirer la l’écran affichera: ou les sondes) Ready For Use (Prêt à La ou les sondes sont désinfectées et l’utilisation) doivent être retirée(s). Étape 15: Press ENTER to Start (Appuyer sur ENTRÉE pour Au retrait, l’écran affichera: commencer) ASTRA TEE (50229_F)