Inspiration time Ti
Resolution
Inspiration/expiration ratio I:E
Ramp rise time Ramp
Resolution
Trigger sensitivity Trigger
Apnoea ventilation Apn. Vent.
Volume guarantee VG
Inspiratory O
concentration FiO
2
Resolution
Performance data
Control principle
Device compliance
Inspiratory resistance
Expiration resistance
Safety valves
Overpresssure
Vacuum
Supply system for patient flow
Max. inspiratory flow
Measurement displays
Inspiratory peak pressure
Pos. end exp. pressure
Mean airway pressure
Range
Resolution
Accuracy
Minute volume MVi, MV
Range
Resolution
Accuracy
T
0...90
Breathing frequency f
Range
Resolution
Accuracy
T
0...90
Instructions for Use Carina SW 3.n
0.3 to 8 seconds
0.1 second
1:39 to 2:1
Auto (device uses the measured breathing rate to calculate
the optimum value)
0.1 to 2.0 seconds
0.1 second
Off, Normal, Sensitive
On, Off
On, Off
21 to 100 %
2
1 %
Time-controlled, volume-controlled and pressure-controlled
<4.5 mbar at 60 L/min; <3.5 mbar at 30 L/min
<6.5 mbar at 60 L/min; <4.0 mbar at 30 L/min
Open the breathing system in the event of a fault
Opens above 60 mbar
Opens below –3 mbar
High dynamic blower
180 L/min at 21 Vol.% O
120 L/min at 100 Vol.% O
PIP
PEEP
P
mean
0 to 80 mbar
1 mbar
±(8 % of the measurement + 1.2 mbar) (with defined hose system
and a bacterial filter)
0 to 40 L/min, BTPS
0.1 L/min
±10 % of the measurement or ±0.6 L/min, the higher value applying
(without leakages)
approx. 35 s
0 to 80 bpm
1 bpm
±1 bpm
approx. 60 s
and normal ambient conditions
2
and normal ambient conditions
2
Technical Data
115